Prelude System Clean¶
A full system cleaning procedure for the Prelude peptide synthesizer.
Before the System Clean¶
-
Wash RVs
- Swap out the glass reaction vials (RVs) for plastic RVs
- Navigate to Operations > Cleaning > Wash RVs
-
Empty all solvent bottles (except DCM)
- Use only the wax-coated bottles
- Make sure each solvent line is connected to a bottle
- Have all amino acid bottles on as well
-
Prime and pressurize the DCM bottle
-
Start the System Clean
- Navigate to Operations > Cleaning > System Clean
After the System Clean¶
-
Dump DCM out of all bottles (amino acid and solvent), except the DCM bottle itself, and rinse with acetone
-
Wash frits and solvent lines (the lines and the small white piece at the end of each line) with acetone